
43
Section 6: Verification
6.1 Verifying Temperature
To verify temperature:
1. Place the temperature probe inside the calibration block. Note
the inside of the PAV after the pan holder is removed.
NOTE: This calibration block is from the ATS PAV Verification Kit.
See Section 8.3.
2. Place the other hole in the calibration block over the RTD on
the bottom of the vessel. Once the block is secure on the
RTD and the temperature probe, slide the PAV lid closed. Use a
small wrench or popsicle stick to keep the lid from pinching
the wire on the temperature probe. DO NOT BOLT THE
LID DOWN.
3. Verify the temperature.
a. Set up the setpoint and start the PAV.