26
Load all ten pans into the sample
1.
holder even if there are not ten
samples to age.
Quickly and carefully return the sample
iv.
holder to the PAV. The vessel temperature
will decrease when the oven and vessel
lids are open. To avoid a long temperature
recovery, attempt to load the PAV and
secure the vessel and oven lids as quickly
and as safely as possible.
NOTE: Do not load samples into the PAV
until the PAV is heated to the setpoint and
the “Preheat Ready” screen is displayed.
Close the vessel lid and secure the bolts
v.
according to the diagram.
Tighten in sequential order with the
vi.
wrench provided. Go around twice to
tighten bolts to 55 inch/pounds.
Pivot Bolt.
DO NOT REMOVE.