MAN - BBR2S - REV: Original | E. Operation
New Button
Creates a new user using default values. Press “Save” once you have created a new user.
Delete Button
Deletes a saved user. Once pressed it will verify that you want to delete the selected user. If you answer “Yes” the
user will be permanently deleted from the system. To prevent software lockout, the system will not allow you delete
the Default user.
E.5 Standardization Process
Before standardization, verify that the bath is set at your desired testing temperature. Set the chiller controller to the
testing temperature desired.
From the Main Screen press the “Standardization” button to enter the Standardization Screen (Figure E.4).
NOTE: The standardization button will not be shown on the main menu unless the currently
logged in user has permission to perform standardization.
The BBR2S’s system status lights, located at the bottom right of the screen, indicate whether or not a component
requires standardization.
•
Red = the component requires standardization
•
Yellow = the component requires verification
•
Green = the component is ready for testing
Prior to standardization, verify that the information entered into the necessary fields matches the information
provided on the Certificate of Conformance that was shipped with the gauge kit being used for the standardization.
Make adjustments as necessary. You will also need a BBR2S gauge kit.
WARNING: Be certain to perform ALL STEPS in the standardization menu for each of the
Figure E.4 - Standardization Screen