Troubleshooting
Error symptom
Possible cause
Remedy
Leakage
Leakage tester is
connected incorrectly
Check connections between
the attachment for pressure
compensation, tube of leakage
tester, and leakage tester
Insertion tube is
defective
Send in the ATMOS Scope for
repair
Protection cap is not
on connector
Push protection cap firmly onto
connector
Deflection lever is
stiff, not working
Defective tip mecha-
nism
Send in the ATMOS Scope for
repair
No image on the PC
Camera USB driver
is not installed on
the PC
Install camera USB driver
ATMOS Scope is not
recognized by USB
interface; the USB
interface may be
turned off
USB port via Windows/control
panel/power settings must be
permanently activated