Maintenance and service
6
Maintenance and service
Maintenance, repairs, and periodic tests may only be carried out by persons who
have the appropriate technical knowledge and who are familiar with the product.
To carry out these measures, the person must have the necessary test devices and
original spare parts.
ATMOS recommends: Work should be carried out by an authorized ATMOS service
partner. This ensures that repairs and testing are carried out professionally, original
spare parts are used, and warranty claims remain unaffected.
6.1
Periodic tests
At least every 24 months a repeat test of the electrical safety should be performed
according to IEC 62353.
6.2
Sending in the device
1. Clean and disinfect the product and accessories according to the operating
instructions.
2. Enclose any used accessories with the product.
3. Use a tightly sealable shipping bag for sending in the endoscope (for example,
the supplied infection control return bag, REF 090.0297.0).
4. Fill in the form QD 434 “Delivery complaint / return shipment” and the respective
decontamination certificate
.