Reprocessing
•
The insertion tube must not bend less than a minimum bending radius of
40 mm during reprocessing.
5.1.4
Further cleaning instructions
Image quality can be reduced even by the slightest contamination such as finger
-
prints on the tip cover glass.
After use, you can protect the ATMOS Scope from dust by storing it on a soft cloth
and covering it.
5.2
Preparing and completing reprocessing
Prior to reprocessing
1. Prior to cleaning, put the protection cap onto the USB connector.
2.
Make sure that the protection cap is fitted securely on the USB connector.
3. Carry out the leakage test before immersing the device in liquids (see chapter
“3.6.2 Performing a leakage test” on page 22). A leaky device will become
damaged by penetrating liquids and can no longer be reprocessed. Send in the
device for servicing.
After reprocessing
Perform a function check:
4. Carry out a visual inspection of the entire ATMOS Scope for visible damage.