Notes on EMC
12.0 Notes on EMC
•
Medical electrical equipment is subject to special precautions with regard to EMC and must be installed
acc. to following EMC notes.
•
Portable and mobile HF communication facilities can influence medical electrical equipment.
•
The use of other accessories, other converters and cables than stated may lead to an increased emission
or a reduced interference immunity of the equipment or system.
Guidelines and Manufacturer´s Declaration - Emissions
The ATMOS
®
E 341 Battery is intended for use in the electromagnetic environment specified below. The
customer or user of the ATMOS
®
E 341 Battery should ensure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions acc.to CISPR 11
Group 1
The ATMOS
®
E 341 Battery uses RF energy
only for its internal function. Therefore, its HF
emissions are very low and it is unlikely that
nearby electronic devices will be affected.
RF Emissions acc. to CISPR 11
Class B
The ATMOS
®
E 341 Battery is suitable for use
in all establishments, including domestic,
and those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions according to
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
according to IEC 61000-3-3
Corresponds
Guidelines and Manufacturer´s Declaration - Immunity for ATMOS
®
E 341 Battery
The ATMOS
®
E 341 Battery is intended for use in the electromagnetic environment specified below. The
customer or user of the ATMOS
®
E 341 Battery should ensure that it is used in such an environment.
Immunity Test
IEC 60601- Test
Level
Compliance Level
Electromagnetic Environment -
Guidance
ESD acc. to IEC
61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood, concrete, or
ceramics tile. If floors are synthetic,
the relative humidity should be at
least 30 %.
Fast electrical
transient/burst IEC
61000-4-4
± 2 kV Mains
± 1 kV I/Os
± 2 kV Mains
Mains power quality should be that
of a typical commercial or hospital
environment.
Surges IEC 61000-4-5 ± 1 kV
common-mode
± 2 kV differential
mode
± 1 kV common-mode
± 2 kV differential
mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
according to IEC
61000-4-11
< 5 % U
T
(> 95 % Dip
of the UT for 0.5
Cycle)
40 % U
T
(60 % Dip of
the UT for 5 Cycles)
70 % U
T
(30 % Dip of
the UT for 25 Cycles)
< 5 % U
T
(> 95 % Dip
of the UT for 5 s)
< 5 % U
T
(> 95 % Dip
of the UT for 0.5
Cycle)
40 % U
T
(60 % Dip of
the UT for 5 Cycles)
70 % U
T
(30% Dip of
the UT for 25 Cycles)
< 5 % U
T
(> 95 % Dip
of the UT for 5 s)
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
ATMOS
®
E 341 Battery demands
continued function even in case of
interruptions of the energy supply,
it is recommended to supply the
ATMOS
®
E 341 Battery from an
uninterruptible current supply or a
battery.
Magnetic field at
power frequency
50/60 Hz acc. to IEC
61000-4-8
3 A/m
Inapplicable
Power frequency magnetic
fields should be that of a
typical commercial or hospital
environment.
NOTE
U
T
is the AC mains voltage prior to application of the test level.