NOTE

: Once you select the start button an audible exit alarm will begin to sound and the 

Sanidyne

®

 

Premium

 will begin 

counting down from 1:30 seconds to allow all inhabitants to leave the area before the UV treatment begins. The words 

“Leave Area” will now appear on the lower status bar of the Hand-held Remote Controller. (See Figure 4, for the specific 

location). The Occupancy Sensors located on the top of the unit (Shown on Figure 1) will begin to operate at this time, 

providing an additional safety area around the unit. If the sensors detect movement close to the 

Sanidyne

®

 

Premium

 

when less than 30 seconds remains on the countdown time, the 

Sanidyne

®

 

Premium

 will automatically reset giving 30 

seconds more to leave the Operation Area. The Countdown time as well as the Treatment Time will all be displayed on the 

Hand-held Remote Controller. See Figure 4, for the specific location.

Occupancy Sensors do not protect the entire Operation Area of the 

Sanidyne

®

 

Premium

. Under no 

circumstances should this sanitizer be permitted to operate with human, or animals present in the Operation Area.

 

4. 

The Hand-held Remote Controller displays the Lamp Status; When the lamp operation indicators on the display glow yellow 

the lamp is on, when the lamp is off it displays grey. (See 

Figure 4

 to locate lamp operation indicators) Once the treatment 

time has expired the unit will turn off the Germicidal Ultraviolet Lamps and it is now safe to return to the area. 

If at any time during the Treatment Time the Occupancy Sensors detect someone near the

 

Sanidyne

®

 

Premium

, the fixture will immediately turn off and send an alarm to the Hand-held Remote Controller and 

the 

Sanidyne

®

 

Premium

. If this event occurs the alarm can be turned off on the remote by pressing anywhere on the 

display or at the 

Sanidyne

®

 

Premium

 

by pressing the START/STOP

 on the Keypad. If the Alarm is turned off at the 

Sanidyne

®

 

Premium

 

the operator can reset the fixture by pressing the 

START/STOP button on the keypad controller 

again to begin the 1:30 countdown to allow all inhabitants to leave the area. After the 1:30 has elapsed the 

Sanidyne

®

 

Premium

 will power on and complete the remaining Treatment Time left prior to the alarm.

If the Hand-held remote is used to stop the alarm, the operator must return to the area and ensure that the area is safe

 

to proceed. Select the START/STOP button on the keypad controller to begin the 1:30 countdown to allow all inhabitants 

to leave the are. After the 1:30 has elapsed the 

Sanidyne

®

 

Premium

 

will power on and complete the remaining

 

Treatment Time left prior the alarm.

s

etting

 d

ate

 

and

 t

ime

Option 1 – Using the tablet’s Settings menu:

1.  Navigate to the tablet’s settings menu by tapping the 

Settings

 icon on the Home Screen or in the App Directory.

2. 

Scroll down to 

Date & Time

 under 

System

.

3.  Use the 

Set Date

Set Time

, and 

Select Time Zone

 functions to set the correct date and time.

Option 2 – If starting in the LampControl app Settings menu:

1.  Press the 

Settings

 Button. (See 

“Hand-Held Remote Controller Overview”

 on 

pages 8–9

.)

2.  Press 

Set Date/Tim

e.

3.  Press 

Admin Pin

. Using keyboard, enter Pin number 1234. (This will bring you to the tablet’s 

Settings

 menu.)

4. 

Scroll down to 

Date & Time

 under 

System

.

5.  Use the 

Set Date

Set Time

, and 

Select Time Zone

 functions to set the correct date and time.

s

aving

 

a

 s

ession

Saving a Session:  At the end of a Session or after a Session is Stopped the 

“Save This Session”

 box will appear.  

To save a Session:

1.  Press 

Yes

 under Save Session.

2.  Press on 

Note Entry

3.  Type in Session Notes, press 

OK

Clearing Total Operating Time or Clearing Session Data:

1.  Press 

Settings

 Button

2.  Press 

Settings

3.  Press on the box next to 

“Reset Total Operating Time”

 and/or 

“Clear Session Data.”

4.  Press 

Home

 Button.

Summary of Contents for Sanidyne PREMIUM

Page 1: ...ctions Save These Instructions UV Ultraviolet Radiation Protect eyes and skin from direct and reflected radiation DANGER Manufacturers Engineers Sales Service Germicidal Ultraviolet Equipment Lamps 37...

Page 2: ...16 MAINTENANCE LOG 17 19 USER ASSISTANCE 20 Patent Notice 20 WARRANTY PRODUCT REGISTRATION 20 These instructions generally describe the installation operation and maintenance of the Sanidyne Premium...

Page 3: ...3 Read and follow all notices and warnings on the sanitizer 14 SAVE THESE INSTRUCTIONS SAFETY WARNINGS All personnel should be alerted to the potential hazards indicated by the product safety labeling...

Page 4: ...or process No warranty or representation is made by the manufacturer with respect to suitability or performance of this equipment or to the results that may be expected from its use The user should p...

Page 5: ...le of Operation The Sanidyne Premium design has been carefully conceived to provide direct germicidal ultraviolet exposure for air and surface disinfection in unoccupied areas to disinfect and prevent...

Page 6: ...nspect all packing materials to prevent the loss of accessories spare parts or instructions Do not discard packing materials until Sanidyne Premium is assembled checked and confirmed The Sanidyne Prem...

Page 7: ...tart Stop Button Begins treatment time countdown also used to restart unit after an intruder is sensed in Operating Area 5 Power Indicator Glows RED when power is applied to unit and will flash during...

Page 8: ...te and 30 seconds to exit area b The words LEAVE AREA appear on the lower status bar c Alarm on unit begins to sound d The word PrEP flashes on the Keypad during the Countdown Time when Treatment Time...

Page 9: ...ler 13 ON OFF Button Turns the power to the device on or off and can be used to restart the device To turn on the device this button must be pushed in and held for 2 seconds If the device is on the Ho...

Page 10: ...ment operation will take place It is the users responsibility to post signs or notices for all personnel and persons in the building or facility so that they be made aware that the Operation Area is o...

Page 11: ...than 30 seconds remain on the countdown time the Sanidyne Premium will automatically reset giving 30 seconds more to leave the Operation Area Occupancy Sensors do not protect the entire Operation Area...

Page 12: ...n the display or at the Sanidyne Premiumby pressing the START STOP on the Keypad If the Alarm is turned off at the Sanidyne Premiumthe operator can reset the fixture by pressing the START STOP button...

Page 13: ...r table etc 2 Measure the longest distance from the Sanidyne Premium to the farthest wall Use this length to compare to the Distance From Fixtures Time Minutes shown on Figure 5 below 3 For example if...

Page 14: ...D AN ULTRAVIOLET BLOCKING FACE SHIELD MUST BE WORN TO COVER AND PROTECT EYES AND EXPOSED SKIN NOTE STER L RAY lamps used in the Sanidyne Premium have ArmorLiteTM Safety Shield which is a protective co...

Page 15: ...Surelite Ballast not operating Disconnect the fixture from utility power then replace ballast Sanidyne Premium fails to shut off Microprocessor assembly faulty Wearing protective clothing and face shi...

Page 16: ...side and can be placed in supplied wall holder Uses 1 AAA battery 6 UV C Dosimeter Card 00 0373 Contains a patented colorimetric indicator that changes color at 25 50 and 100 mJ cm2 to visibly confirm...

Page 17: ...17 MAINTENANCE LOG Date Maintenance Performed Service Person Comment s...

Page 18: ...18 MAINTENANCE LOG Date Maintenance Performed Service Person Comment s...

Page 19: ...19 MAINTENANCE LOG Date Maintenance Performed Service Person Comment s...

Page 20: ...including warranties of merchantability and fitness for a particular purpose whether or not purposes or specifications are described herein We further disclaim any responsibility whatsoever to the cus...

Reviews: