Manual, F/T Sensor, Serial Axia
Document #9610-05-Serial Axia-05
Pinnacle Park • 1031 Goodworth Drive • Apex, NC 27539 • Tel:+1 919.772.0115 • Fax:+1 919.772.8259 •
25
3.5 Pin and Wire Assignments for Serial and Power Connection
CAUTION:
Ensure the cable shield is properly grounded. Improper shielding on the
cables can cause communication errors and an inoperative Serial Axia sensor.
The following section provides the pin assignment for the connector on the Serial Axia sensor and
applicable connectors on the cables from
. Refer to the following table or
for supply voltage ratings. Refer to
Section 9.4—Cable Specifications
for additional cable
technical specifications.
Table 3.3—Power Supply
1
Power Source
Voltage
Power Consumption
Minimum
Nominal
Maximum
Maximum
DC Power
12 V
24 V
30 V
1.5 W
Notes:
1. The power supply input is reverse polarity protected. If the power and ground to the power supply
inputs are plugged in reverse, then the reverse polarity protection stops the incorrectly wired supply
input from damaging or powering on the sensor.
3.5.1 Pin Assignment for the 6-pin M8 Male Connector on the Sensor
The following table details the signals and corresponding pin numbers on the M8 connector (ZC28)
on the Serial Axia sensor.
Table 3.4—Pin Assignment for the 6-pin M8 Male Serial Axia Sensor Connector
Connector Schematic
Pin Number
Signal
1
2
6
3
4
5
1
RS485+
2
RS485-
3
CLK_SYNC
4
Ground (for CLK_SYNC)
5
V +
6
V -
3.5.2 Pin Assignment for Cable P/N 9105-CP-ZC38-U-1.8
The following table details the signals and corresponding pin numbers for unterminated wires on
cable P/N 9105-CP-ZC38-U-1.8 that connects from the USB cable P/N 9105-C-ZC33-ZC34-XXX
to the customer’s device.
Table 3.5—Cable P/N 9105-CP-ZC38-U-1.8, Wire Jacket Color, and Signal
Wire Jacket Color
Signal
-
Shield
Black
Ground
Red
CLK_SYNC