PiezoClamping
Calibration Kit
Page 9 of 14
6.
Configuration and use
This section describes how to configure and use the PiezoClamping Calibration Kit to
calibrate the PiezoClamping
equipment.
Step-by-step procedure:
Step 01
Turn the PiezoClamping
equipment on. Then press [
▲
] or [
▼
] to access
the charge display screen, as shown in the figure below:
Step 02
Connect the Calibration Kit connector
“
OUT
”
[4] to the PiezoClamping
panel using the BNC-BNC (A) cable.
Step 03
Switch the Calibration Kit ON using the toggle switch ON/OFF [1], one of
the charge indicator LEDs [5] should light up. Then ensure the toggle switch
OUT/CHARGE is in
“
CHARGE
”
position.