PiezoClamping
Calibration Kit
Page 13 of 14
8.
Troubleshooting
Problem
Possible cause
Solution
The PiezoClamping
Calibration Kit does not
turn on.
The ON/OFF toggle switch is
on the “
OFF
” position.
Move the toggle switch
to the “
ON
” position.
The battery is depleted.
Charge the battery (see
topic 7).
PiezoClamping cannot
do the charge
measurement, or
The measurements are
not reproducible.
The BNC-BNC cable is not
connected.
Connect the BNC-BNC
cable.
The BNC-BNC cable
presents poor contact or
one of the internal wires
has been interrupted.
Replace the BNC-BNC
cable.
The PiezoClamping
Calibration Kit is faulty.
Send the unit for
service.
The PiezoClamping
Calibration battery does
not charge.
The USB cable is not
connected.
Connect the USB cable.
The USB cable presents
poor contact or one of the
internal wires has been
interrupted.
Replace the USB cable.
The smartphone charger is
faulty.
Replace the smartphone
charger.