PiezoClamping
Calibration Kit
Page 10 of 14
Step 04
Press the charge selector [2] to select the charge (5.50, 11.0, 16.50 or
22.00 µC).
Step 05
Press [START] on PiezoClamping
to tare the equipment and start a new
charge measurement.
Step 06
Change the toggle switch OUT/CHARGE to
“
OUT
”
in order to discharge the
selected charge in the PiezoClamping
.
Step 07
Read the result on the PiezoClamping
display, the result should match the
selected charge with ±5% tolerance.
To perform additional charge measurements, repeat steps 04 to 07.