Page 11 of 14
Table 2
–
Measurement acceptance intervals.
Selection Nominal charge Q Minimum Maximum
1
5.50 µC ±5 %
5,22 µC
5.77 µC
2
11.00 µC ±5 %
10.45 µC
11.55 µC
3
16.50 µC ±5 %
15.67 µC
17.32 µC
4
22.00 µC ±5 %
20.90 µC
23.10 µC
As the values measured by PiezoClamping are within Table 2 acceptance ranges, the
PiezoClamping
equipment is calibrated and no action is needed. If any the results
were out of range, the PiezoClamping
must be sent for service.
Turn the Calibration Kit off after use it to avoid battery depletion.