- 5 -
2. Basic Operations
CAUTION
●
Operate away from direct sunlight at a stable temperature with as little fluctuation
as possible.
●
Acclimate both the instrument and sample solution to room temperature to ensure
accurate results. Measurements may fluctuate with hot or cold samples. Wait to
press
F
after placing the sample on the sample stage. Alternatively, press
F
multiple times.
●
Operate on a flat work surface.
●
The displayed temperature is that of the sample stage and may not necessarily
match the temperature of the sample.
1. Zero-setting
This is the operation to set the calibration point (zero).
m
Zero-set the instrument at the beginning of each day before use.
m
Regular zero-setting throughout the day is recommended to ensure accurate results.
①Press
I
to turn it on.
m
Zero setting can be done in any
scales.
②Wipe the sample stage clean.
③Fill the sample stage with tap water up to
the fill line.
N
Use room temperature tap water. The
instrument should also be acclimated to
room temperature prior to operations.
④Press
G
.
000
flashes and stops.
Fill line
(Approx. 3mL)