![ATAGO 5050 Instruction Manual Download Page 4](http://html.mh-extra.com/html/atago/5050/5050_instruction-manual_2995799004.webp)
Multi-scale: Angle of Rotation; Specific Rotation; Concentration
Main Features
Press
H
to select the desired scale
–
Specific Rotation (factory default) ; or
Concentration
The One and Only Polarimeter That Doesn
’
t Use Observation Tubes
The instrument measures angle of rotation equivalent to a 20mm observation tube and
displays angle of rotation converted to 100mm observation tube and 200mm observation
tube.
Measurement Evaluation Indicator
The
measurement
values
are
evaluated
against
the
pre-set
tolerance,
and
the green/red
light will indicate the results.
See page 10 for how to configure upper and lower limits.
Red light
= Outside tolerance
Green light
= Within tolerance
Outside
tolerance
Red light
Green light
Red light
Within
tolerance
Outside
tolerance
Upper limit
Lower limit
Angle of Rotation
(Conversion to 100mm
observation tube )
Angle of Rotation
(Conversion to 200mm
observation tube )
Angle of Rotation
(Conversion to 100mm
optical path length)
Angle of Rotation
(Conversion to 200mm
optical path length)