b. When the
Options
pop-up menu is displayed,
is highlighted. To print the LQC
results, press the right soft key.
c. To hide the
Options
pop-up menu, use the
key to highlight
Hide Options
and
press the right soft key.
d. To return to the
Main Menu
, press the left soft key.
13. To run the next liquid control (high or low) after viewing or printing the LQC results, press
the right soft key to display the
Options
pop-up menu. Use the
key to highlight
Next
Control
and press the right soft key. Repeat steps 4–12.
Testing a Patient Sample
Assay-specific test cartridges that are uniquely compatible with the ASTUTE140
®
Meter are used
to test patient samples. The test cartridges approved for use with the ASTUTE140
®
Meter should
not be used with any other testing device. Similarly, only test cartridges approved for use with the
ASTUTE140
®
Meter can be used. For a list of approved tests for use with the ASTUTE140
®
Meter
and to purchase assay-specific test cartridges, go online or consult your sales representative.
Each new test lot must be registered prior to use of any Test cartridges from that lot (See “Test
Lot Registration” on p. 52 for instructions).
To test a patient sample, perform the following steps:
1. Press the
key to display the
Main Menu
(if running patient sample immediately after
successful log in, the
Main Menu
will automatically be displayed).
2. When the
Main Menu
is displayed, use the navigation keys to highlight the
Run Patient
icon. Press the right soft key to select
Run Patient
.
Figure 65.
Main Menu screen
with Run Patient icon
highlighted