ASTUTE140
®
Meter
ASTUTE140
®
Meter Operation
©Astute Medical, Inc. 2021 PN 300005 Rev G 2021/03/03
54
10. When prompted, hold the Test RFID Card for the test lot next to or against the numeric
keypad to register the test lot information and select
OK
by pressing the right soft key.
11. If registered correctly, a screen displaying the test lot number, the test type and the
analytes detected by the test will appear. Press the right soft key to select
Accept
. The test
lot that was just registered should now appear in the list of registered test lots.
12. If registered incorrectly, an error message will appear. Press the right soft key to select
OK
and close the error message. Repeat steps 10–11. If registered incorrectly a second time,
contact Technical Support (See “Contacting Technical Support ” on p. 7 for contact
information).
13. After use, place the Test RFID Card in its sleeve and return it to the kit from which it was
removed. Once all the cartridges in the kit have been used, the Test RFID Card and the
Test cartridge kit may be discarded in accordance with local regulations.
14. To register another test lot, locate the Test RFID Card for the test lot to be registered and
repeat steps 7–13.
Liquid Control Lot Registration
High and low liquid controls must be run to verify the precision and accuracy of the assay
procedure, including the performance of the Test cartridges and assay-specific reagents.
Each assay-specific liquid control kit includes a high and a low liquid control. Each liquid control in
the kit must be registered prior to first use. Operators may register the liquid controls only if the
Supervisor granted registration permission in the default settings; otherwise, Supervisors must
register the liquid controls.
Figure 59.
Register Test Lot
screen