Quality Control Settings
Perform the following steps to set the time intervals at which LQC and EQC must be run. It is
recommended that LQC be run every 30 days (as well as with every new lot of Test cartridges,
with each new shipment of Test Kits, or in accordance with the best practices of the institution).
EQC should be run in accordance with the best practices of the institution.
1. Press the
key to display the
Main Menu
.
2. Use the navigation keys to highlight the
Supervisor Menu
icon and press the right soft key.
3. When the
Supervisor Menu
is displayed,
General Meter Settings
is highlighted. Press
the right soft key to display the
General Settings
screen.
4. When the
General Settings
screen is displayed,
Default Settings
is highlighted. Press the
right soft key to display the
Default Settings
screen.
5. On the
Default Settings
screen, use the
key to highlight
QC Settings
and press the
right soft key
.
Figure 24.
Main Menu screen
with Supervisor
Menu icon
highlighted
Figure 25.
Default Settings
screen with QC
Settings highlighted