AsCARD B5 & B56
When
INOP
is signalised, the ECG leads are printed as straight lines. It does not matter
which electrode is badly contacted. If the INOP is displayed the auto examination will not
be carried out. Check the electrodes contact before the examination start.
Choose the record speed
Choose the record sensitivity
Choose the lead to record
Choose registration mode
Press the
key. The display shows:
The unit stores the ECG in the memory simultaneously from all 12 leads. After
memorising 10 sec course, the unit starts the ECG printout (described before the
examination start). The length of ECG leads is described in the unit configuration.
The examination can be interrupted by pressing the
key.
If in configuration parameters the auto examination with analysis was set or during
data collecting the
key was pressed, the unit starts to calculate the ECG
parameters. The display shows:
After 20 sec. the unit starts to print the report including:
- P-QRS-T complex of the most numerous class with marking beginnings and ends
of segments. The lower line shows the class size. The record speed is always
50mm/s.
- table with the values of segments amplitude in all leads
- written interpretation of P-QRS-T complexs calculated parameters.
After finishing the printout the unit goes back to the beginning position:
NOTE!
The measurement results are reliable if on the QRS class representative record the
beginnings and ends of segments are marked correctly and the number of detected