AsCARD B5 & B56
Choose registration mode
(AsCARD B5),
(AsCARD B56).
Press the
key. The unit starts electrocardiogram registration - the chosen
ECG leads, speed, sensitivity and filters are presented on the display.
During the electrocardiogram registration it is possible to change such parameters
as:
- ECG leads
- record speed
- record sensitivity
- the filters switching on or off
- record mode
Every parameters change is automatically described on the display and registered
on the electrocardiogram.
ECG registration lasts until the
key is pressed - then registration is broken
off.
Press the
key to switch off the unit.
The unit switches off automatically. If the unit is battery operated and no keys are
pressed or nothing is printed, it switches off after 3 minutes.
To switch the unit on press the
key.
4.2 How to perform automatic examination.
Switch on the unit with the use of the
key.
The display shows:
The unit supervises the electrodes contact with patients body and signalises poor
contact with
INOP
,