SECTION 7 DIAGNOSTICS AND TROUBLESHOOTING
________________________________________________________________________
7-5
Status Messages: Possible Causes:
Corrective Actions:
Sensor Disconnected
[0014]
1.
Disconnected sensor.
2.
Poor or contaminated
connection between
sensor and PIC.
3.
Disconnected PIC.
4.
Defective PIC.
5.
Defective BISx.
1.
Connect the sensor.
2.
Connect/clean connection
between sensor and PIC.
3.
Connect the PIC.
4.
Replace the PIC.
5.
Replace BISx Bulkhead
Connector or replace BISx.
Last Sensor Check
Failed [0016]
1.
At least one element of
sensor has too high
impedance, and EXIT
pressed (before sensor
check completes).
2.
Poor sensor connection
3.
Incorrect sensor
application.
4.
Defective PIC.
5.
Defective BISx.
1.
Verify Sensor Check
passes.
2.
Check sensor connection.
3.
Read Instructions on
sensor package and re-
apply sensor.
4.
Replace the PIC.
5.
Replace BISx Bulkhead
Connector or replace BISx.
Unrecoverable BISx
Error [0024]
1.
Poor communication
between BISx host cable
and monitor.
2.
Defective BISx.
3.
Defective monitor.
1.
Unplug BISx from
monitor. Power down
monitor. Power up
monitor, then reconnect
BISx. If error returns
contact Technical Service.
2.
Replace the BISx.
3.
Replace the monitor.