SECTION 1 SAFETY PRECAUTIONS
________________________________________________________________________
1-2
USE ONLY THE POWER CORD SUPPLIED BY THE MANUFACTURER.
NEVER ADAPT THE PLUG FROM THE MONITOR TO FIT A NON-
STANDARD OUTLET.
U.S.A. REQUIREMENT: FOR PROPER GROUNDING, THE POWER
RECEPTACLE MUST BE A THREE-WIRE GROUNDED OUTLET. A
HOSPITAL GRADE OUTLET IS REQUIRED. NEVER ADAPT THE THREE-
PRONG PLUG FROM THE MONITOR TO FIT A TWO-SLOT OUTLET. IF
THE OUTLET HAS ONLY TWO SLOTS, MAKE SURE THAT IT IS
REPLACED WITH A THREE-SLOT GROUNDED OUTLET BEFORE
ATTEMPTING TO OPERATE THE MONITOR.
IF THE INTEGRITY OF THE EXTERNAL PROTECTIVE EARTH GROUND
IS IN DOUBT, THE BIS VIEW SHALL BE OPERATED FROM ITS INTERNAL
BATTERY POWER SOURCE ONLY.
THE CONDUCTIVE PARTS OF ELECTRODES OR SENSOR AND
CONNECTORS, INCLUDING THE NEUTRAL ELECTRODE, SHOULD NOT
CONTACT OTHER CONDUCTIVE PARTS, INCLUDING EARTH.
TO REDUCE THE HAZARD OF BURNS IN THE HIGH-FREQUENCY
SURGICAL NEUTRAL ELECTRODE CONNECTION, THE SENSOR OR
ELECTRODES SHOULD NOT BE LOCATED BETWEEN THE SURGICAL
SITE AND THE ELECTRO-SURGICAL UNIT RETURN ELECTRODE.
TO REDUCE THE HAZARD OF BURNS DURING USE OF BRAIN-
STIMULATING DEVICES (e.g., TRANSCRANIAL ELECTRICAL MOTOR
EVOKED POTENTIAL), PLACE STIMULATING ELECTRODES AS FAR AS
POSSIBLE FROM THE BIS™ SENSOR AND MAKE CERTAIN THAT SENSOR
IS PLACED ACCORDING TO PACKAGE INSTRUCTIONS.
THE SENSOR MUST NOT BE LOCATED BETWEEN DEFIBRILLATOR
PADS WHEN A DEFIBRILLATOR IS USED ON A PATIENT CONNECTED
TO THE BIS VIEW SYSTEM.
TO MINIMIZE THE RISK OF PATIENT STRANGULATION, THE PATIENT
INTERFACE CABLE (PIC) MUST BE CAREFULLY PLACED AND SECURED.
WHENEVER AN EVENT SUCH AS SPILLAGE OF BLOOD OR SOLUTIONS
OCCURS, RE-TEST GROUND LEAKAGE CURRENT BEFORE FURTHER
USE.
BE SURE THE MONITOR IS MOUNTED SECURELY IN PLACE TO AVOID
PERSONAL INJURY.
MONITOR IS NOT DESIGNED FOR USE IN MRI ENVIRONMENT.