SECTION 10 APPENDIX I
_______________________________________________________________________
10-5
5.
Connect this Test Sensor to the PIC. All impedance tests should complete
successfully, with low impedance values. Typical values using this alternative Test
Sensor are less than 5 K ohms.
The following chart shows expected Test Sensor values:
Test Sensor Values
Electrode # Typical
Range
1
1 K ohm
1-2 K ohms
2
1 K ohm
1-3 K ohms
4
1 K ohm
1-2 K ohms
3
1 K ohm
1-2 K ohms