SECTION 7 DIAGNOSTICS AND TROUBLESHOOTING
________________________________________________________________________
7-4
7.3
BIS VIEW System Messages and Corrective
Actions
When an alarm is triggered, a message appears in the Message Region of the screen. Possible
messages and the recommended operator actions are listed below:
BIS VIEW Messages and Operator Actions
Status Messages: Possible Causes:
Corrective Actions:
None (screen is blank)
Monitor has not completed
its power ON diagnostics
tests.
1.
Monitor may take up to
three minutes to initialize
trend memory. Do not
disconnect equipment or
press keys during this time.
2.
If several minutes have
passed, restart monitor.
3.
Reset monitor.
4.
Replace monitor.
Connect BISx
1.
BISx disconnected.
2.
Defective BISx cable.
3.
Defective BISx.
4.
Defective monitor.
1.
Connect BISx. Verify all
cable connections.
2.
Replace Monitor Interface
Cable.
3.
Replace the BISx.
4.
Replace monitor.
Re-prep Sensor [0013]
1.
Sensor is not fully in
contact with patient’s
skin.
2.
Incorrect sensor
application.
3.
Poor sensor
connections.
4.
Sensor Check fails
5.
Defective PIC.
6.
Defective BISx.
1.
Press electrodes.
2.
Read Instructions on
sensor package to ensure
correct sensor placement.
3.
Check sensor connections.
4.
Re-prep again or replace
sensor. Verify Sensor
Check passes.
5.
Replace the PIC.
6.
Replace BISx Bulkhead
Connector or replace BISx.