aks-S4
Page 34
z4200064_GA_Bett_S4_151015_EN
Technical Data
21
Technical Data
Classification
Class I medical product according to Appendix IX of the MDD
Lying area nominal dimensions:
90 x 200 cm
External dimensions:
108 x 205 cm
Height adjustment:
41 to 91 cm
Back part adjustment:
0° to 70°
Thigh part:
0° to 37°
Lower leg part:
for knee bend 0° to 23°, leg high level up to 14°
Side grill height:
36 cm
Side grill height extender
(accessory)
Side grill height:
55 cm
Castor diameter:
12.5 cm
Safe working load:
170 kg (including patient, mattress and accessories)
max. patient weight
135 kg
Empty weight:
121.5 kg
Separate weights:
< 33.5 kg
Uplifter:
6.4 kg
Climatic conditions:
‑Ambient temperature from 0 °C to 40 °C
‑ Relative humidity from 20% to 80%
‑ Air pressure from 700 hPa to 1060 hPa
‑ Normally composed atmospheric air
Materials used:
Steel (powder coated or galvanised),
Commercially available plastics (POM, ABS, PP, PVC, PA6.6), Rubber,
Wood (varnished or covered with decorative foil).
MDF plates covered with decorative foil
Aluminium (anodised)
Electrical Data
Input:
100 ‑ 240 V
; 50/60 Hz; 90 W
Output SMPS:
29 V ; 2.0 A
Duty cycle:
ED 2 min / Pause 18 min
Protection class:
II
Protection type:
IPX4
Battery for emergency lowering:
9 Volt block battery (alkaline manganese block battery Type 6LR61)
Sound pressure level:
< 53 dB(A) at distance of 1 m
The following requirements for the
electrical system are complied with:
RoHS, REACH and WEEE
All parts and data are subject to constant further development and can thus be different from the
information shown in this document.