General Warnings, Training and Safety Notices – Read This First
NanoScope Manual
DFU-0298-1, Revision 0
Page 5 of 38
system to comply with the requirements of IEC 60601-1. If in doubt, contact a
qualified biomedical technician or your local representative.
18.
The NanoScope Console is provided with a standard VESA MIS-D, 100 mm x 100
mm (3.9 in. x 3.9 in.) threaded hole mounting pattern on the back of the unit
intended for vertical mounting. Use M4 mounting fasteners. The blind thread
depth is 10 mm (0.39 in.). Use the appropriate length screws for the bracket
thickness. This mounting configuration and tip load is qualified for all VESA
mounts rated for at least 5 kg (11 lbs.).
19.
This equipment has been tested and found to comply with the Class A limits for
medical devices to the IEC 60601-1:2005+A1:2012(E) and IEC 60601-1-2:2014.
These limits provide reasonable protection against harmful interference in a
typical medical installation.
20.
This equipment generates and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause harmful
interference to other device(s) in the vicinity.
21.
Interference may occur in any installation environment. To test whether the
NanoScope Console, or other devices in the environment are causing
interference, turn the equipment off and on and identify the cause. Correct the
interference by using one or more of the following measures:
•
Reorient or relocate the receiving device.
•
Increase the separation between devices.
•
Connect the equipment into an outlet on a separate electrical circuit.
•
Consult the manufacturer or field service technician for help.
22.
This unit was not evaluated for use with electrosurgical devices.
23.
After each use, thoroughly clean the NanoScope Console unit with a clean cloth
dampened with commercially available surfactants or surface disinfectants only.
See Cleaning and Disinfecting,
24.
NEVER allow the Console receptacles to have any contact with liquids. If there is
dust or moisture on the receptacles, remove with dry compressed air. ONLY dry
connectors should be plugged into the Console.
25.
NEVER use liquid to clean the accessory device connector contacts. Remove dust
regularly with dry compressed air.
26.
Liquid on the cable connector of the accessory device can damage the device.
Before connecting the cable, ensure the receptacles are clean and dry.
After use, gather the Handpiece sterile packaging, the Handpiece, and all Handpiece
components and dispose of them in the appropriate biohazard containers. Observe all local,
state and national waste management regulations. Do not dispose of Waste Electrical and
Electronic Equipment (WEEE) as unsorted municipal waste.