Technical Specifications
NanoScope Manual
DFU-0298-1, Revision 0
Page 16 of 38
3.2
Ambient Conditions for Operation
Table 6
NanoScope System Ambient Conditions for Operation
Temperature
10° to 30° C (50° to 86° F)
Relative humidity
30% to 75%
Air pressure
700 hPa to 1060 hPa (21 in. Hg to 31.3 in. Hg)
3.3
Ambient Conditions for Storage (in shipping packaging)
Table 7
NanoScope System Ambient Conditions for Storage
Temperature- Console
-20° C to 50° C (-4° F to 122° F) up to 1 month;
-20° C to 45° C (-4° F to 113° F) up to 3 months;
-20° C to 20° C (-4° F to 68° F) up to 12 months
Temperature- Handpiece
-40° C to 50° C (-40° F to 122° F) up to 1 year
Relative humidity
10% to 90%, non-condensing
Air Pressure
500 hPa to 1060 hPa (15 in. Hg to 31.3 in. Hg)