Ventilation
The
TECSource
has vent holes on the rear and left side of the unit. You must
not block these vent holes, or overheating may occur, causing damage to the
unit.
Rack Mounting
Rack mounting kits (
p/n 1401-RM-1
or
p/n 1401-RM-2
) for standard 19” racks
are available for the
TECSource
, and supports the rack mount of one (
1401-
RM-1
) or two (
1401-RM-2
) units in a 2U (3.5”) high opening. Because the unit
draws air from the side, and therefore inside the rack housing, be sure that the
internal rack ambient temperature (which will typically be several degrees higher
than room ambient) does not exceed the unit’s operating temperature.
Warm-up and Environmental Considerations
In order to achieve the highest level of accuracy, the
TECSource
should be
powered on for at least one hour prior to taking measurements. In addition,
ensure that the unit is not operating outside the ambient temperature range or
humidity conditions.
CAUTION
Do not operate the unit above +40°C ambient, and
ensure the instrument is properly ventilated, or the unit
may overheat and possible damage to the instrument
may occur.