Page 16 · 5300 Series
TECSource
User’s Manual
Menu
Description
Factory
Default
AutoTune PID
AutoTune PID Function
See using the AutoTune Function later in this
manual for details on this item.
Sensor Coeffs
Sensor Coefficients Menu
ThermA
The A term in the Steinhart-Hart equation.
1.12924E-03
ThermB
The B term in the Steinhart-Hart equation.
2.34108E-04
ThermC
The C term in the Steinhart-Hart equation.
0.87755E-07
RTD A
The A term in the RTD equation.
3.98480E-03
RTD B
The B term in the RTD equation.
-0.58700E-06
RTD C
The C term in the RTD equation.
4.00000E-12
RTD R
0
The
R
0
term (in
Ω
) in the RTD equation.
100.00
AD590 M
The slope term in the AD590 correction
equation.
1.00000
AD590 B
The offset term, in °C, in the AD590 correction
equation.
0.000
LM335 M
The slope term in the LM335 correction
equation.
1.00000
LM335 B
The offset term, in °C, in the LM335 correction
equation.
0.000
Sens Cal M
The slope term for user sensor calibration of
RTD or thermistor measurements.
1.0000
Sens Cal B
The offset term for user sensor calibration of
RTD or thermistor measurements.
0.00
Ω
Comm Menu
Communications Menu
Baud
This sets the baud rate for the RS232 serial
port. See the
Computer Interfacing Manual
which is included on the CD that accompanied
this product.
9600
Err While
Rmt
To turn off the display of errors while in remote
mode, set this value to “No”. To display errors
while in remote mode, set this value to “Yes”.
Yes