Page 14 · 5300 Series
TECSource
User’s Manual
Menu
Description
Factory
Default
PID I
The integral term of the PID loop. Will only be
available if
Gain
is set to
PID
.
0.01
PID D
The derivative term of the PID loop. Will only be
available if
Gain
is set to
PID
.
0
Sensor
Sets the temperature measurement sensor.
Can be
Therm 100uA
,
Therm 10uA
,
Therm,
LM335
,
AD590
,
RTD
, or
RTD (4-wire)
.
Therm 100uA
T-Low Lim
T-Low Lim
is the lower temperature limit. If
operating in T Mode and the actual
temperature drops below this value, the output
will be turned off. Temperature limits not
monitored in R Mode or Ite Mode. Limit
monitoring can also be disabled in software
(see TEC:ENAB:OUTOFF command).
-99°C
T-High Lim
T-High Lim
is the upper temperature limit. If
operating in T Mode and the actual
temperature rises above this value, the output
will be turned off. Temperature limits not
monitored in R Mode or Ite Mode. Limit
monitoring can also be disabled in software
(see TEC:ENAB:OUTOFF command).
125°C
R-Low Lim
R-Low Lim
is the sensor low limit. If operating
in R Mode and the actual sensor measurement
drops below this value, the output will be
turned off. R limits are not monitored in Ite
Mode. Limit monitoring can also be disabled in
software (see TEC:ENAB:OUTOFF command).
0.01k
Ω
R-High Lim
R-High Lim
is the sensor high limit. If operating
in R Mode and the actual sensor measurement
exceeds this value, the output will be turned off.
R limits are not monitored in Ite Mode. Limit
monitoring can also be disabled in software
(see TEC:ENAB:OUTOFF command).
45k
Ω
Tol Time
Tolerance time is the amount of time, in
seconds, that the actual temperature must be
within the set point tempe/- the Tol
Temp value for the unit to be considered in
tolerance.
5 seconds
Summary of Contents for TECSource 5305
Page 1: ......
Page 35: ...5300 Series TECSource User s Manual Page 35...
Page 43: ...5300 Series TECSource User s Manual Page 43 NOTES...