background image

Page 6 · 

586

 

TECPak

 User’s Manual 

 

 

 

Quick Start 

 

 
The 

586

 

TECPak

 was designed with ease of use in mind, and you will likely 

have little need for this manual for almost all of the features the unit offers. This 
section will show how you can quickly get the unit up and running in almost no 
time. 
 
The 

TECPak

 has a universal input power supply, accepting 90 to 240 VAC, and 

50 or 60 Hz. This covers all conventional power worldwide, but ensure your AC 
power meets these requirements.  
 
Plug the AC cord into the unit and into the wall outlet. Turn on the power switch 
located on the IPC, and the unit will power up. The AC Power LED should light, 
and the On / Error LED should remain dark. 

 
Install the 

Arroyo Control

 software located on the CD that came with the 

instrument (it can also be downloaded from our web site), located in the 

Software

 folder on the CD. Once the software is installed, plug the instrument 

into the USB interface. The instrument may also ask for drivers, which are also 

found on the CD. 
 
Launch 

Arroyo Control

, and in the TECSource panel (if one is not visible, click 

Add Panel in the lower right of the window, and click Add a TEC Panel), select 
the communications port to which your 

TECPak

 is connected. This will usually 

be the highest numbered port in the list. Click Connect, and 

Arroyo Control

 will 

begin controlling the 

TECPak

Once connected, you will be able to adjust 

settings, limits, and turn the output on or off. 
 
Next, connect the cable between your LaserMount or other fixture and the 

Output

 connector of the 

TECPak

. We recommend using our cables as they 

have been designed to work well with the 

TECPak

. If using your own cables, 

ensure they have been properly wired according to the pin-out of the 

TECPak

 

and your fixture. 
 
Finally, set the set point to an appropriate temperature and turn the output on. 

You will see the voltage and current begin driving the fixture to the set point you 
have chosen. Depending on the thermal size of your fixture, it may take seconds 
or several minutes to reach the set point.  
 
If you notice the temperature is oscillating around the set point and not 
stabilizing, you may need to adjust the 

Gain

 setting in the 

Settings…

 dialog. 

You can use the AutoTune feature to automatically calculate the best PID values, 
or select from a set of eight factory preset values that typically cover most 
applications. To use the factory gains, if the temperature is quickly jumping up 
and down, the 

Gain

 will typically need to be reduced. If the temperature is slowly 

moving up and down, try a higher 

Gain

. You may need to experiment with 

Summary of Contents for TECPak 586 Series

Page 1: ......

Page 2: ...19 Using the Current Limit 19 Using the Voltage Limit 19 Understanding the Sensor Inputs 20 Applying Sensor Limits 20 Selecting the Fixture 21 Working With Thermistors 22 Working With RTDs 23 User Cal...

Page 3: ...and actual temperatures What s in the Box Along with the 586 TECPak itself a CD with electronic copies of this manual the Computer Interfacing Manual and USB drivers are included For USA customers a...

Page 4: ...the NOTES heading are not safety related and are intended simply to point out important information If at any time any of the following conditions exist or are suspected of existing discontinue use o...

Page 5: ...oid electrical shock ensure a 3 prong power cord is used and is plugged into a earth grounded receptacle Failure to do so can result in severe injury or death CAUTION There are no user serviceable par...

Page 6: ...your TECPak is connected This will usually be the highest numbered port in the list Click Connect and Arroyo Control will begin controlling the TECPak Once connected you will be able to adjust setting...

Page 7: ...ower supply accepting 90 to 240 VAC and 50 or 60 Hz This covers all conventional power worldwide but ensure your AC power meets these requirements Powering Up the Unit Connect the AC power cord to the...

Page 8: ...ir gap is required to the sides of the unit to ensure proper air flow No air gapping is required above or below the units and they can be rack mounted immediately above other equipment no rack spacers...

Page 9: ...nsor equation and directly select the sensor set point This can be useful when only the desired sensor value is known or when the sensor to temperature conversion values are not available for your sen...

Page 10: ...Specify the mount connected to the TECPak If using a LaserMount select the appropriate model Otherwise select User Defined User Defined TEC MOUNT I Lim I Lim sets the current limit of the temperature...

Page 11: ...Mode Limit monitoring can also be disabled in software see TEC ENAB OUTOFF command 99 C TEC LIM TLO T High Lim T High Lim is the upper temperature limit If operating in T Mode and the actual temperatu...

Page 12: ...seconds TEC TOL Tol Temp Tolerance temperature is a temperature band in C around the set point temperature When the actual temperature is within this band for longer than the Tol Time setting then th...

Page 13: ...nectors in ohms This setting allows for accurate voltage measurement at the TEC by removing the voltage loss of the cable 0 0080 TEC CABLER T Rate Selects a desired temperature ramp rate in degrees Ce...

Page 14: ...ating control of the TECPak allowing you to change limits settings and set points as well as monitor the operation of the instrument You can install Arroyo Control from the CD or download it from our...

Page 15: ...TE command While not recommended as it can lead to confusion when the controller is used later in properly wired setups this allows for correct operation of system when the TE pins are wired backwards...

Page 16: ...e wizard appears insert the CD you received with the TECPak and follow the on screen instructions Once the drivers are installed to determine the COM port number go to Control Panel and select System...

Page 17: ...sed Output On turn on the output when circuit is closed You can also invert the logic level of the input using the same command For example if the function was inverted and set to Output On the output...

Page 18: ...ent to be wired with a series resistor and must use the parallel resistor option When using a series resistor the higher the resistor value the dimmer the LED Do not use too high of a resistance value...

Page 19: ...utput the current will not exceed the limit at any time in any mode Set the limit by using the TEC LIM ITE command Operating at the current limit is common especially when transitioning from one tempe...

Page 20: ...f your device or mount The two sensor inputs support Sensor 1 supports thermistor LM335 RTD and 4 wire RTD sensors Sensor 2 supports thermistor and RTD sensors One or both can be enabled for monitorin...

Page 21: ...s disabled which means only a temperature limit of the active sensor will cause the temperature to shut down Once a determination is made that a temperature limit condition exists the settings in the...

Page 22: ...nce of the thermistor drops very quickly In the typical control range 0 C to 70 C typical 10K thermistors offer good sensitivity to changes in temperature and this is the range in which most 10K therm...

Page 23: ...the coefficients for the BetaTHERM 10K3A1 thermistor A 1 12924x10 3 B 2 34108x10 4 C 0 87755x10 7 You can change the coefficients using the TEC CONST command or through the Settings dialog within Arro...

Page 24: ...urement modes a conventional two wire measurement mode which is subject to this error and a four wire measurement mode that uses separate sensor and source lines to remotely sense the actual resistanc...

Page 25: ...ading offset The offset is kilo ohms k for thermistors and ohms k for RTDs Controlling the Temperature Rate of Change Some applications require that temperature is changed at a specific rate to preven...

Page 26: ...Output Where is the error in the system expressed as Target Actual The controller can calculate ideal PID values using the AutoTune function discussed in detail in the next section To manually adjust...

Page 27: ...tely put the instrument into AutoTune mode The TEC AUTOTUNE query can be used to monitor the process of AutoTune A response of 1 indicates AutoTune is in progress Once complete the instrument will res...

Page 28: ...s on 3 Similar to mode 1 the fan is turned on whenever the TEC output is on but when the TEC is turned off the fan will remaining running for an addition number of minutes as defined by the delay sett...

Page 29: ...alculating cable resistance is to short the TE and TE connections making sure whatever you are using to short the terminals is not itself adding resistance to the circuit and is rated for the current...

Page 30: ...to the TEC module as is reasonable To enable the remote TEC voltage sense function send a TEC VENSE 1 The reported TEC voltage will now be measured through pins 2 and 9 instead of TEC and TEC at the i...

Page 31: ...24hr C 5 0 01 Measurement Channels Current Resolution mA 10 Accuracy mA 10 20 20 30 Voltage Resolution mV 10 Accuracy V 0 05 Sensor6 10 A Thermistor Range k 0 1 450 Resolution k 0 01 Primary Accuracy...

Page 32: ...0 3 1 RTD Range 20 192 Resolution 0 01 Accuracy reading 0 03 0 1 Current Limit Resolution A 0 1 Accuracy A 0 2 Other Fan Supply 8 12V 350mA max TEC Connector Female 17W2 Computer Interface USB 2 0 Po...

Page 33: ...ue that was outside the allowable range USB Serial only E 202 Invalid data type When trying to parse the message the data was in an invalid format USB Serial only E 204 Suffix not valid An invalid num...

Page 34: ...no longer in calibration mode E 419 TEC not stable The TEC is considered stable if the temperature has changed less than 0 02 C for more than 20 seconds E 433 Not a TEC The TEC CHAN command attempted...

Page 35: ...586 TECPak User s Manual Page 35 Mechanical Drawings Front View Rear View Bottom View...

Page 36: ...Page 36 586 TECPak User s Manual Right Side View Top View Left Side View...

Page 37: ...d there are no user serviceable fuses Service Service and repair for the 586 TECPak can be obtained by contacting the distributor from where you purchased the instrument or directly from Arroyo Instru...

Page 38: ...epairs will be made and the instrument returned freight prepaid Repaired products are warranted for the balance of the original warranty period or 90 days whichever first occurs Limitation of Warranty...

Page 39: ...scharge 4kV contact 8kV air EN 61000 4 3 2010 Radiated Immunity 3V m EN 61000 4 4 2012 Electrical Fast Transients Burst 1kV AC 0 5kV I O EN 61000 4 5 2014 Surges 0 5kV differential mode 1kV common mod...

Page 40: ...Page 40 586 TECPak User s Manual Copyright 2019 Arroyo Instruments All Rights Reserved P N 530 1051 Rev A...

Reviews: