586
TECPak
User’s Manual · Page 13
Name
Description
Factory
Default
Computer
Command
Ext Fan
Mode
Controls when the fan operates.
See the
External Fan Control
section
below for more
information.
Auto
TEC:FAN
Ext Fan Off
When
Ext Fan Mode
is set to
Delay
, this setting defines the
number of minutes to delay
turning off the fan after the TEC
output has been turned off. See
the
External Fan Control
section
below for more information.
5 minutes
TEC:FAN
Cable R
The resistance of the cable and
connectors, in ohms. This
setting allows for accurate
voltage measurement at the
TEC by removing the voltage
loss of the cable.
0.0080
Ω
TEC:CABLER
T Rate
Selects a desired temperature
ramp rate in degrees Celsius
per minute. Set to 0.0°C/min to
disable rate limiting.
0.0°C/min
TEC:TRATE
Sensor Settings
ThermA
A term in thermistor equation.
1.12924E-03 TEC:CONST
ThermB
B term in thermistor equation.
2.34108E-04 TEC:CONST
ThermC
C term in thermistor equation.
0.87755E-07 TEC:CONST
RTD A
A term in the RTD equation.
3.98480E-03 TEC:CONST
RTD B
B term in the RTD equation.
-0.58700E-06 TEC:CONST
RTD C
C term in the RTD equation.
4.00000E-12 TEC:CONST
RTD R
0
R
0
term in the RTD equation.
100.00 TEC:CONST