586
TECPak
User’s Manual · Page 11
Name
Description
Factory
Default
Computer
Command
Sensor
Sets the temperature
measurement sensor. Can be
Therm 100uA
,
Therm 10uA
,
Therm,
LM335,
RTD
, or
RTD (4-
wire)
.
Therm
100uA
TEC:SENS
Active
Sensor
Selects sensor 1 or sensor 2 for
feedback control.
1
TEC:ACTIVESENS
T-Low Lim
T-Low Lim
is the lower
temperature limit. If operating in
T Mode and the actual
temperature drops below this
value, the output will be turned
off. Temperature limits not
monitored in R Mode or Ite
Mode. Limit monitoring can
also be disabled in software
(see TEC:ENAB:OUTOFF
command).
-99°C
TEC:LIM:TLO
T-High Lim
T-High Lim
is the upper
temperature limit. If operating in
T Mode and the actual
temperature rises above this
value, the output will be turned
off. Temperature limits not
monitored in R Mode or Ite
Mode. Limit monitoring can
also be disabled in software
(see TEC:ENAB:OUTOFF
command).
125°C
TEC:LIM:THI
R-Low Lim
R-Low Lim
is the sensor low
limit. If operating in R Mode and
the actual sensor measurement
drops below this value, the
output will be turned off. R
limits are not monitored in Ite
Mode. Limit monitoring can
also be disabled in software
(see TEC:ENAB:OUTOFF
command).
0.01k
Ω
TEC:LIM:RLO