Page 12 ·
586
TECPak
User’s Manual
Name
Description
Factory
Default
Computer
Command
R-High Lim
R-High Lim
is the sensor high
limit. If operating in R Mode and
the actual sensor measurement
exceeds this value, the output
will be turned off. R limits are
not monitored in Ite Mode. Limit
monitoring can also be
disabled in software (see
TEC:ENAB:OUTOFF
command).
45k
Ω
TEC:LIM:RHI
Tol Time
Tolerance time is the amount of
time, in seconds, that the actual
temperature must be within the
set point tempe/- the
Tol Temp value for the unit to
be considered in tolerance.
5 seconds
TEC:TOL
Tol Temp
Tolerance temperature is a
temperature band (in °C)
around the set point
temperature. When the actual
temperature is within this band
for longer than the Tol Time
setting, then the unit is
considered to be in tolerance.
0.1°C
TEC:TOL
H/C Mode
This selects the heating and/or
cooling mode of the
TECPak
.
See the section below titled
“Resistive Heaters and
Heat/Cool Only Modes” for
more information.
Heat/Cool
TEC:HEATCOOL
Ext Fan
This selects the voltage for the
auxiliary fan power supply. See
the
External Fan Control
section
below for more information.
Off
TEC:FAN
Ext Fan Pwr When
Ext Fan
is set to
Custom
,
this sets the specific fan voltage
from 4.0 to 12.0 volts. See the
External Fan Control
section
below for more information.
12.0
TEC:FAN