2-Wire versus 4-Wire Measurements
One concern in using RTDs are their relatively low resistance (typically 100
Ω
at
0°C), and small
Ω/
°C. Because of these two factors, the resistance of the cable
used to connect to the sensor can become a significant error in the sensor
measurement. Most Arroyo Instruments controllers offer two RTD measurement
modes: a conventional two wire measurement mode, which is subject to this
error, and a four wire measurement mode that uses separate sensor and source
lines to remotely sense the actual resistance of the RTD and eliminate the cable
or connector resistances.
When using 4-wire measurement mode, you must select ‘RTD (4-wire)’ as the
sensor type, and then connect the and Remote at one side of
the RTD, and Sensor– and Remote Sensor– to the other side of the RTD. Make
these connections as close to the sensor as possible.
The drawings below illustrate how 2-wire and 4-wire connections work. Note that
4-wire measurements require all four wires to be brought through the cable to
the mount. The
1262 TECSource
cable carries this connection through to the
mount, but the
1260B
cable does not.
RTD 2-wire Measurement
RTD 4-wire Measurement