NOTE
Arroyo Instruments offers TEC cables designed to connect directly
between our
TECSource
products. If you use your own cables, ensure
the connections are properly made between the instrument and the
mount, and that proper grounding techniques are used. The pin-out of
the connectors can be found later in this document.
Connecting the water supply:
The
274
comes with factory installed brass barbs suitable for 3/8” I.D. hose, but
they can be replaced with any 1/8” NPT male fitting. It is recommended that
when using the barbs, pipe clamps be used to secure the hose to the
274
.
NOTE
Earth Grounding Considerations
If your application requires that the cold plate be earth grounded, you
should directly wire the cold plate to earth ground.
When not using the 274 for prolonged periods of time, it is recommended that
the water be drained from the unit and allowed to air dry. A burst of compressed
air is also beneficial to blow out any residual water inside the unit.
Mounting Plate
The
274
TECMount
has a bread board configuration with M3 holes on 10mm
centers. Optional blank adapter plates are available in several sizes (3.2”, 4.2”,
and 5.2” square) to allow for custom mounting patterns or larger mounting
surfaces. See mechanical information below.