3-2
5.
Press both manual switches to raise unit to highest position.
Manual switches are located on lower foot-end of unit frame, as
shown at right.
6.
Remove shipping blocks:
• Remove foot-end shipping block:
-
Pull pin from top of block.
-
Rotate block 90° toward caster frame.
-
Insert pin into top of block to lock block in place.
• Remove head-end transport block:
-
Pull pin from top of block.
-
Slide block toward head end of unit.
-
Rotate block 180°.
-
Insert pin into top of block to lock block in place.
7.
Press
Yes
button. The
RotoProne Calibration Utility
screen will appear.
The
No
button may be pressed to skip
calibration and continue directly to the
Welcome To RotoProne / New Patient
screen.
However, it is recommended that calibration
be performed any time the RotoProne Unit
is powered up after transport, especially if it
has traveled significant distance or over door
thresholds.
Do not calibrate scales
. Scale
calibration should be performed by trained
ArjoHuntleigh personnel.
8.
Press
Start Calibration
button. Unit calibration will begin.
• Patient surface will raise to maximum height, move to a slight Trendelenberg angle and lower back
down.
• Once patient surface has stopped moving, calibration status indicator will change to a green
calibrated indicator and calibration is complete.
Unit temperature may affect scale calibration. To ensure accuracy of scale calibration, do not
immediately zero scales when delivered from extreme temperature environments.
Manual Switch Door
Calibrate Scales