2-10
1.
Press the
Hold Patient Weight
button; the button will change to green and the scale freezes.
2.
Add or remove items from the patient surface; wait approximately five to ten seconds.
3.
Press the
Hold Patient Weight
button again; the button will change to red and the scale resumes
function with no change in weight.
Zero scale
Use Zero to recalibrate the Scale to zero kilograms (or zero pounds) prior to patient placement, with any
linens, equipment, etc. already placed on patient surface or frame.
• Press
Zero
button; a confirmation screen will appear.
• Press
Yes
button. A
Zeroing Scales
screen will appear.
DO NOT TOUCH UNIT UNTIL SCREEN CHANGES
.
Current Weight
display will change to zero and the
Scale
screen will reappear.
Display
Use Display to change the display to kilograms or pounds. Default scale display is in kilograms.
• Press the
Display
button; the
Current Weight
display will change from kilograms to pounds.
Weigh Delay
Use Weigh Delay to postpone weighing for a specified amount of time while tubes, equipment, etc. are lifted
and then to hold the resulting weight value until it is read and recorded. Weigh Delay adds a weight reading
to the Weight Trend Chart at the end of the specified time.
1.
Press
Weigh Delay
button;
Weigh Delay
screen will
appear.
2.
Set desired delay time. Use the up and down arrow
buttons to increase and decrease the weigh delay time.
3.
Press
Start
button.
4.
Add or remove items from patient surface.
5.
As soon as countdown is complete, an alert beep will
sound. The current weight is saved in Weight Trends
and the scale resumes regular function.
6.
Press
Done
button;
Scale
screen will reappear.