2-5
Checklist
The short
Checklist
screen appears prior to initiation of supine therapy or when rotating the patient from
prone to supine. It requires assessment of the patient’s invasive lines and tubes, intubation system and
placement in the head support.
The long
Checklist
screen appears prior to rotating the patient from supine to prone. In addition to assessing
patient’s invasive lines and tubes, intubation system and head placement, it requires assessment of the
patient’s arm and abdominal support.
For detailed information on equipment check and assessment of patient position, refer to the
Starting
Therapy
and
Prone Therapy
sections of the
Patient Placement
chapter of this manual.
1.
Assess all invasive lines and tubes. If secure, press
Check Tubing
button. The arrow will change to a
check mark.
2.
Examine patient’s artificial airway system. If secure, press
Check Airway
button. The arrow will change
to a check mark.
3.
Examine patient’s placement in head support assembly. If secure, press
Check Head Support
button.
The arrow will change to a check mark.
4.
Examine placement of patient’s arms in arm supports. If secure, press
Check Arm Slings
button (long
checklist only). The arrow will change to a check mark.
5.
Examine placement of chest and pelvic packs. If secured and positioned properly, press
Check Abdomen
Support
button (long checklist only). The arrow will change to a check mark.