3-1
PaTIenT PlaCeMenT
It is recommended that all chapters of this manual be reviewed prior to product use.
Carefully read the
Contraindications, Risks and Precautions and Safety Information sections in the Introduction
chapter of this manual prior to placing a patient on the RotoProne Therapy System
.
Preparation for Patient Placement
1.
Lock caster brakes by pressing red pedal on each caster until clicked in place.
2.
Configure patient surface, as required (add all packs, accessories, equipment to unit).
Any equipment on the unit frame is included in the weight displayed. Packs and all equipment (i.e.
Foley bags) must be present on the patient surface to ensure accurate scale calibration.
3.
Verify the following:
• Ensure patient surface is at 0° supine position.
• Ensure lock pin is engaged.
4.
Plug power cord into a properly grounded wall outlet to
activate power to the
RotoProne
Unit. Press on / off
switch. The software will boot and the
Set Date
/
Time
screen will appear on the main display.
Do not use a wall outlet controlled by a wall
switch.
• Press up and down arrows to set a new date and
time.
• Press
24 Hour Time
button to change time to 24-
hour format, if desired.
• Press
Save New Time
button to continue.
or
• Press
Time is Correct
button to continue without
saving a new date and time.
• The
Calibration
screen will appear.