Electromagnetic Compatibility
42
(continued)
Rated maximum
output power of
transmitter
W
Separation distances according to frequency of transmitter m
Recommended Separation Distance Between Portable and Mobile RF Communications
Equipment and the Maxi Sky
1000
- for Equipment and Systems that are not
Life-Supporting
Recommended separation distances between portable and mobile RF communications equipment and the
Maxi Sky
1000
.
The Maxi Sky
1000
is intended for use in electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Maxi Sky
1000
can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communication equipment (transmitters) and the
Maxi Sky
1000
as recommended below, according to the maximum output power of the communications equipment.
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.24
0.1
0.38
0.38
0.74
1
1.2
1.2
2.4
10
3.8
3.8
7.4
100
12
12
24
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in
metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
d
3.5
V
1
-------
P
=
d
3.5
E
1
-------
P
=
d
7
E
1
-------
P
=