7
This example shows the pump running and the left
garment inflated to 40 mmHg.
Refer to “Battery Operation” on page 12 when
operating the pump under battery power.
TO STOP press and hold the green RUN
button a
second time. The green power indicators will go out and
therapy will stop.
Recommended
Checks:
• Check display symbols to confirm that the
correct type of garment has been connected.
• During garment inflation, check the display to
confirm that the correct pressure is being applied
and that there are no fault messages appearing.
• Check that there are no kinks in the pump hoses
and garment tubing.
• Check that the pump hoses and connectors do not
cause the patient any discomfort.
• Regularly check that the garments remain
correctly fitted to the patient.
To Use One Garment
When the pump is in standby (see “Standby Mode” on
page 5) connect one garment to either pump hose
connector.
The display will confirm that only one garment is
connected.
If a garment is disconnected while the pump is
running, the system will alarm
Lo
. If the single
garment option is still required, press the RUN
button to reset the system and press RUN
again to restart therapy.
If, at any time, a second garment is connected while
the pump is running, the pump will automatically
change to two garment operation.