6
Garment Application
Use the correct garment type as ordered by the patient’s
physician.
For detailed instructions on correct fitting and use,
carefully read and follow the instruction provided with
every garment set.
Connect the garments to the pump hose set connectors
and ensure the connectors click into place.
Check that the connection and garment type are
confirmed correctly on the display.
Typical garment options and therapy parameters
TO START, press and hold the green RUN
button
on the pump. The green power indicators will illuminate
and therapy will begin.
Garment Type/
LCD Display
Inflation Pressure
(mmHg)
Inflation Hold
(seconds)
Cycle (seconds)
Foot garments
e.g., FG100-200
130
3
30
Single chamber
garments
e.g., DVT10-40,60
40
12½
60