13
When running under battery power, a battery symbol
appears in the top right hand corner of the display.
In order to maximize battery life, the power alarm
indicators will not illuminate when operating the
Flowtron Universal
under battery power. You may
verify therapy is, or is not, being provided via the
LCD screen.
The amount of charge remaining in the battery pack is
indicated by the black block inside the battery symbol.
When the battery has only 5% of the charge remaining,
the battery symbol will flash and the remaining charge
will be displayed as a percentage. Recharge the battery
as soon as possible.
Then, when the battery has 1% charge, the audible alarm
will sound and the pump will stop running. Recharge the
battery.
Battery Storage and
Disposal
If the device is not to be used for an extended period,
charge (refer to “Recharging the Battery Pack” on
page 15), remove and store the battery pack. If the
battery pack is to be put into long term storage it should
be recharged at least once every three to six months.
A faulty battery pack should be sent back to
ArjoHuntleigh for recycling or correct disposal.