52
Electromagnetic compatibility
Medical electrical equipment needs special precautions regarding EMC (Electromagnetic compatibility) and needs to
be used according to the EMC information below. The
Carevo
is intended for use in the electromagnetic environment
specified below. The customer or the user of the
Carevo
should assure that it is used in such an environment.
WARNING
Stacking or placing other electrical equipment next to this device is not recommended, it can
interfere with the equipment's operation and safety. Portable and mobile radio-frequency (RF)
communications equipment can interfere with this equipment's operation and safety.
WARNING
The equipment may cause radio interference or may disrupt the operation of nearby
equipment. It may be necessary to take actions, such as reorienting, relocating the equipment
or shielding the location.
Guidance and manufacturer's declaration - electromagnetic emissions
Emissions test
Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
Carevo
uses RF energy only for its internal function. Therefore, its RF emis-
sions are very low and are not likely to cause any interference in nearby elec-
tronic equipment.
RF emissions
CISPR 11
Class B
Carevo
is suitable for use in all establishments, those directly connected to the
public low-voltage power supply network that supplies buildings used for
domestic
purposes.
EN 61000-4-8
3A/m at 50/
60 Hz
Magnetic Immunity: If functional disturbance occurs on the
Carevo
it may be necessary to move the
Carevo
further from the source
of power frequency magnetic field. The power frequency magnetic field
should be measured in the intended use locations to assure that it is
sufficiently low.
Guidance and manufacturer's declaration - electromagnetic immunity
Emissions test
IEC 60601 test level
Compliance
level
Electromagnetic environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Summary of Contents for Carevo
Page 1: ... with people in mind Carevo 04 BA 08_9US CA September 2014 INSTRUCTIONS FOR USE ...
Page 11: ...11 3 8 7 9 23 15 1 2 6 12 21 8 18 25 17 19 4 5 24 10 22 2 20 11 13 16 14 26 ...
Page 31: ...31 Intentionally left blank ...
Page 33: ...33 Fig 1 ...
Page 37: ...37 Intentionally left blank ...
Page 51: ...51 Intentionally left blank ...