35
Always Follow these 24 Steps for Proper
Cleaning and Disinfection.
If an ArjoHuntleigh shower panels will be used, see respective
IFU
.
Mechanical cleaning (Step 1-8 of 24)
1
Put on protective gloves and eye wear.
(See Fig. 1)
2
Make sure the disinfectant is mixed according to the
instructions on the disinfectant bottle label or according to
the mix ratio stated in the shower panel
IFU
.
(See Fig. 2)
3
Remove the lower drainage and the pillow.
4
Rinse the
Carevo
with water. Start from the top of the
mattress down towards the castors.
(See Fig. 3)
5
Scrub the
Carevo
with disinfectant using a soft bristle
cleaning brush to remove contamination. Start from the top
of the mattress moving down towards the castors.
6
Use a bottle brush soaked in disinfectant to clean the upper
drainage.
7
Remember to clean the underside of mattress, support plates
and outside of the upper drainage.
8
Using a soft bristle cleaning brush, scrub the pillow and the
lower drainage separately. Avoid contact with the mattress
until the pillow has been disinfected.
continue with the steps on the next page
Fig. 1
Fig. 2
Fig. 3
Summary of Contents for Carevo