6. PRIOMA BED SPECIFICATIONS
Dimensions
Overall Length
219.5 cm
Overall Length Base Extension Extended (19.5cm)
239.0 cm
Bed Extension Length
19.5 cm
Linen Rack (Bed Stripper) Extension
30.0 cm
Mattress Surface Length - Between Panel Faces
203.0 cm
Mattress Surface Length - Extended
222.5 cm
Overall Width
100.0 cm
Deck Height Range - Tente Twin 125mm Castors
37.0cm to 82.0cm
Trendelenburg Tilt
0 to 14º
Reverse Trendelenburg
0 to 14º
Backrest Angle Adjustment
0 to 70º
Knee Break Angle Adjustment
0 to 35º
Calf Positioning Angle Adjustment
0 down to 25º
Mattress Size Dimensions
Recommended Mattress Size
W=87 cm L=198 cm
Mattress Thickness Minimum
12.5 cm
Mattress Thickness Maximum
17.0 cm
Product Weight
(approximate)
Without Side Rails Fitted
128 Kg
With Side Rails Fitted
142 Kg
Safe Working Load SWL Safety Factor = 2 x SWL
250 Kg
Maximum Patient Weight
185 Kg
Electrical Data (230V)
Power In
3 A max. At 230V ac 50Hz
Duty Rating
Intermittent 10%, Max 2 min / 18 min
Electric Shock Protection
Class 1 Type B
Liquid Ingress Protection
IP54
Internally Powered Equipment
(Battery Backup Only)
2 x 12V Series Connected Batteries
Sealed, Rechargeable Lead/Acid Gel
1.2 amp hrs.
Standards
The Prioma acute care bed has been built to comply with International Standard
IEC 60601-1 Section 9 and IEC 60601-2-52 Section 9 (Edition 3.1 2012-08)