4. ELECTRONIC FUNCTIONS DESCRIBED
- Continued
CAUTION
When Emergency Patient Positioning Functions are activated, care should be
taken over the potential of the beds moving parts to either trap or crush the
patient, caregivers and or other equipment.
Pressing the SINGLE BUTTON PRESS button, the bed will automatically level all
positioning functions back to a flat deck position whilst lowering the bed to a safe
working height.
(9) Emergency Trendelenburg (head down tilt) Function
NOTE:
IF FITTED Emergency Trendelenburg Positioning Function is available
on the Attendant Control Panel (ACP), Attendant Control OpenBus Mini Panel
(ACOM) and the Inset Nurse Side Control Panels at the head end of the bed.
Note:
For the Emergency Patient Positioning Functions It is
NOT
necessary to
press the ‘up/down’ buttons as the operation is programmed to operate directly
from the associated button.
Pressing the
“TRENDELENBURG”
button will lower the head and raise the foot
of the bed to the maximum tilt of 14°, while simultaneously lowering the back rest
and knee break back to a flat deck position –
NO
pause if passing through
horizontal.
(10) Lock Out Functions
NOTE:
IF FITTED the lock out function is available on the Attendant Control
Panel (ACP), Attendant Control OpenBus Mini Panel (ACOM) ) and the Inset
Nurse Side Control Panels at the head end of the bed.
This facility allows users to selectively disable some of the electrical features of
the bed. In normal operation the key functions of the bed should be in an
(unlocked) state of use. This is indicated on the ACC panel by LED lights above
the function buttons.
To lock and unlock each function hold the “key” or “lock” button and at the same
time press the function button that is required to be locked or unlocked. An LED
light will illuminate when the function is in the locked position.
(11) Battery Charge Indicator - Battery Backup
NOTE:
IF FITTED the battery charge indicator is available on the Attendant
Control Panel (ACP), Attendant Control OpenBus Mini Panel (ACOM) ) and the
Inset Nurse Side Control Panels at the head end of the bed.