5. GENERAL BED OPERATION
-
Continued
5.6 Manual Cardiopulmonary Resuscitation (CPR) Function
A clearly marked emergency CPR release handle is available on both sides of the bed.
The manual release function returns the beds backrest to a level position in the event of an
emergency resuscitation being required.
Emergency Lowering:
To lower the backrest in an emergency, pull up on the red CPR
release handle
(1)
. The handles are positioned at the side and towards the head of the
bed. The lowering movement is assisted by a gas damper, but
CAUTION
is required as
the backrest may lower quickly under patient weight.
In addition to the CPR lever and
IF FITTED
there are also emergency positioning functions
(CPR) on the Attendant Control Panel (ACP), Attendant Control OpenBus Mini panel
(ACOM) and inset Nurse Control Panel at the head end rails of the bed. These buttons will
drive the bed deck into a low flat horizontal position automatically, regardless of the beds
configuration before starting the movement.
WARNING
The backrest can fall quickly - keep hands clear to avoid trapping.