Page | 7
Commercial in confidence. User Manual Revision CTK-SM-REV8-060720
2.
DECONTAMINATION
WARNING
•
Disconnect the bed from the mains power supply before carrying out
cleaning procedures.
•
Do not allow the mains plug or power supply cord to get wet when
cleaning the bed.
•
Disconnect the bed from the mains power supply before carrying
out cleaning procedures.
•
The following procedure also applies to the beds accessories
but does not include the mattress.
2.1 Cleaning
Clean all surfaces with a disposable cloth soaked in a neutral detergent and a comfortable
hot temperature for bare hands.
Mattress platform sheet tops and head/foot panels can be removed from the bed and
cleaned separately.
Start by cleaning the upper sections of the bed work along all horizontal surfaces. Work
methodically towards the lower sections of the bed and clean the castors last. Take extra
care to clean areas that may trap dust or dirt.
Wipe over with a new disposable cloth moistened with clean water and dry with disposable
paper towels. Allow the cleaned parts to dry before replacing the mattress.
2.2 Disinfecting
After cleaning the bed as described above, wipe all surfaces with sodium
Dichloroisocyanurate (NaDCC) at a concentration of 1,000 parts per million (0.1%) of
available chlorine.
In the case of pooling body fluids, e.g. blood, the concentration of NaDCC should be
increased to 10,000 parts per million (1%) of available chlorine.
CAUTION
•
Do not use abrasive compounds or pads or phenol-based disinfectant
solutions.
•
Do not use jet stream cleaning or wash tunnels.
•
Take care not to remove lubricant from the actuator pistons.